Sensitivity of film measured off-axis ratios to film calibration curve using radiochromic film
DOI:
https://doi.org/10.29384/rbfm.2011.v5.n1.p21-24Abstract
Off-axis ratios of conical beams generated with a stereotactic radiosurgery-dedicated LINAC were measured with EBT2 film and stereotactic diode. The sensitivity of both full width at half maximum (FWHM) and penumbras (80-20% and 90-10%, respectively), with respect to the characteristics of the film calibration curve fit, was investigated. In all cases, penumbras resulted to be more sensitive than FWHM. However, these differences were, in general, smaller than the ones found between EBT2 reference values and the stereotactic diode measurements. The larger variation in OAR parameters was found to depend on whether the fit intersected or not the origin. A 1D gamma-index analysis showed this difference can be important in all measured conical beams.Downloads
Downloads
How to Cite
Issue
Section
License
The submission of original articles to the Brazilian Journal of Medical Physics implies the transfer, by the authors, of the rights of print and digital publication. Copyright for published articles remains with the author, with journal rights on first publication. Authors may only use the same results in other publications by clearly indicating this journal as the original publisher. As we are an open access journal, free use of articles in educational, scientific, non-commercial applications is allowed, as long as the source is cited.