Optimization of the scan protocol in the measurements of coronary artery calcium

Authors

  • Larissa C. G. Oliveira
  • Ilan Gottlieb
  • Fabrício M. de Carvalho
  • Larissa C. Pinheiro
  • Simone Kodlulovich
  • Fernando A. Mecca
  • Ricardo T. Lopes

DOI:

https://doi.org/10.29384/rbfm.2011.v5.n1.p63-66

Abstract

The aim of this study was to evaluate the influence of the tube current applied for studies of calcium score. The research was carried out in a private clinic of Rio de Janeiro, using a 64-slice MDCT scanner and an anthropomorphic cardiac computed tomography phantom. In all images, the Agatston score, the volume and mass of the calcifications, and the noise for each current tube were determined. The average computed tomography attenuation number obtained for all tube currents was 261.6±3.2 HU for the CaHA density insert and -0.2 HU±2.0 for the water insert. The images obtained at lower tube currents were noisier and grainier than those obtained at higher tube currents. However, no significant differences were found in the calcium measurements, which suggest a high potential of patient dose reduction, around 50%, without compromising diagnostic information.

Downloads

Download data is not yet available.

How to Cite

Oliveira, L. C. G., Gottlieb, I., de Carvalho, F. M., Pinheiro, L. C., Kodlulovich, S., Mecca, F. A., & Lopes, R. T. (2015). Optimization of the scan protocol in the measurements of coronary artery calcium. Brazilian Journal of Medical Physics, 5(1), 63–66. https://doi.org/10.29384/rbfm.2011.v5.n1.p63-66

Issue

Section

Artigo Original