Detector characterization of a digital mammography system in 2D and 3D acquisition modes

Authors

DOI:

https://doi.org/10.29384/rbfm.2023.v17.19849001609

Keywords:

radiology, 2D mammography, tomosynthesis, image quality, ; detector characterization

Abstract

The 2D and 3D acquisition modes of the GEHC Senographe Pristina digital mammography system were evaluated through physical figures of merit:  the modulation transfer function (MTF), normalized noise power spectrum (NNPS), quantum detection efficiency (DQE) and image system noise analysis through component decomposition. The response functions of the detector were linear for both acquisition modes regardless of the spectrum used by the equipment (26 kV Mo/Mo and 34 kV Rh/Ag), noting that the gain of the detector in 3D mode is greater to compensate the lower detector air kerma (DAK) per projection. The dominance ranges of each type of noise calculated through the noise components are presented. The quantum component was the dominant one in the range of DAK used, being approximately 80% the total variance at 100 µGy. MTF curves were obtained in 2D and 3D modes in the horizontal and vertical directions, with higher average values obtained from MTF50% and MTF  5mm-1 in 3D versus 2D mode. In 3D mode, MTF was obtained    at distances 20 mm, 40 mm and 50 mm over the detector. A decrease in MTF has been observed with increasing height. In both modes the radial NNPS was always higher for the 26 kV Mo/Mo spectrum than for the 34 kV Rh/Ag spectrum. The same behavior was observed in DQE and this was explained by the smallest amount of photons/mm2µGy for the Mo/Mo spectrum and the obtained NNPS. The values of DQE 0.5 mm-1 to approximately 100 µGy were always greater than 0.5, with higher DQE for 3D mode. These results have shown good performance of the detector used for clinical purposes.

Downloads

Download data is not yet available.

References

Van Engen RE, Bosmans H, Bouwman RW, Dance DR, Heid P, Lazzari B. Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems. Netherlands; 2018.

Van Engen R, Young K, Bosmans H, Thijssen M. European protocol for the quality control of the physical and technical aspects of mammography screening. Part B: Digital mammography. 2006.

Medical electrical equipment - Characteristics of digital X-ray imaging devices - Part 1-2: Determination of the detective quantum efficiency - Detectors used in mammography. IEC. 2007.

Marshall NW, Monnin P, Bosmans H, Bochud FO, Verdun FR. Image quality assessment in digital mammography: part I. Technical characterization of the systems. Phys Med Biol [Internet]. 2011;56(14):4201–20. Disponible en: http://dx.doi.org/10.1088/0031-9155/56/14/002.

Technical Evaluation of GE Healthcare Senographe Pristina digital Breast Tomosynthesis System. National Health System. 2019.

Technical Evaluation of GE Healthcare Senographe Pristina digital Mamography System in 2D. National Health System. 2019.

Ortenzia O, Rossi R, Bertolini M, Nitrosi A, Ghetti C. Physical characterisation of four different commercial digital breast tomosynthesis systems. Radiat Prot Dosimetry [Internet]. 2018;181(3):277–89. Disponible en: http://dx.doi.org/10.1093/rpd/ncy024.

General Electric. Senograph Pristina Operator Manual 5762780EN Revisión 9. 2018.

Monnin P, Bosmans H, Verdun FR, Marshall NW. Comparison of the polynomial model against explicit measurements of noise components for different mammography systems. Phys Med Biol [Internet]. 2014;59(19):5741–61. Disponible en: http://dx.doi.org/10.1088/0031-9155/59/19/5741.

Samei E, Flynn MJ, Reimann DA. A method for measuring the presampled MTF of digital radiographic systems using an edge test device. Med Phys [Internet]. 1998;25(1):102–13. Disponible en: http://dx.doi.org/10.1118/1.598165.

Donini B, Rivetti S, Lanconelli N, Bertolini M. Free software for performing physical analysis of systems for digital radiography and mammography: Free software for physical analysis in digital radiography. Med Phys [Internet]. 2014;41(5):051903. Disponible en: http://dx.doi.org/10.1118/1.4870955.

Boone JM, Fewell TR, Jennings RJ. Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography. Med Phys [Internet]. 1997;24(12):1863–74. Disponible en: http://dx.doi.org/10.1118/1.598100

Marshall NW. Detective quantum efficiency measured as a function of energy for two full-field digital mammography systems. Phys Med Biol [Internet]. 2009;54(9):2845–61. Disponible en: http://dx.doi.org/10.1088/0031-9155/54/9/017

De Física Médica S, Española ;., De Protección Radiológica S. Protocolo Español de Control de Calidad en Radiodiagnóstico. Madrid: SEFM; 2011.

Monnin P, Marshall NW, Bosmans H, Bochud FO, Verdun FR. Image quality assessment in digital mammography: part II. NPWE as a validated alternative for contrast detail analysis. Phys Med Biol [Internet]. 2011;56(14):4221–38. Disponible en: http://dx.doi.org/10.1088/0031-9155/56/14/00

Koutalonis M, Delis H, Pascoal A, Spyrou G, Costaridou L, Panayiotakis G. Can electronic zoom replace magnification in mammography? A comparative Monte Carlo study. Br J Radiol [Internet]. 2010;83(991):569–77. Disponible en: http://dx.doi.org/10.1259/bjr/21753020

Gehealthcare.com. [citado el 1 de febrero de 2023]. Disponible en: https://landing1.gehealthcare.com/rs/005-SHS-767/images/Senographe%20Pristina%20Datasheet.pdf

Published

2023-03-28

How to Cite

Frederico Alvarez, M. J., & Banguero, Y. . (2023). Detector characterization of a digital mammography system in 2D and 3D acquisition modes. Brazilian Journal of Medical Physics, 17, 609. https://doi.org/10.29384/rbfm.2023.v17.19849001609

Issue

Section

Artigo Original

Most read articles by the same author(s)