Study of CT Acquisition Protocols Using Two Head Phantoms
DOI:
https://doi.org/10.29384/rbfm.2023.v17.19849001706Keywords:
Computed Tomography, Phantom, Dosimetry, Image QualityAbstract
Computed Tomography (CT) scans promote a higher dose deposition than conventional radiology exams. These exams have significantly increased patient and collective doses and have become a global public health concern. There is a great need to improve protocols to seek for lower doses while maintaining the diagnostic image quality. The development of phantoms allows the testing of different acquisition protocols. In this study were tested two cylindrical head phantoms of polymethylmethacrylate (PMMA). One CT head phantom is the head standard test with 16 cm in diameter and the other head phantom developed is smaller at 12 cm in diameter. Both phantoms are 15 cm long. Different acquisition protocols were performed on a Philips CT scanner, Access model with 16 channels. The central slice of the phantoms was irradiated successively and measurements were performed using a pencil ionization chamber to obtain the CT air Kerma indexes in PMMA (Ck,PMMA,100) and CT dose indexes (CTDI). From these results, the CT Dose Index values weighted and volumetric (CTDIw, CTDIvol) were obtained to 10 cm scans of the central region of the head phantoms, in helical mode. The scans were performed using different voltage values (80, 100 and 120 kV) and charge (mA.s). Dose values varied from 5.59 to 21.51 mGy. The highest recorded dose value was 21.51 mGy for the smaller head phantom and 19.25 mGy for the standard head phantom with 120 kV. Considering the generation of images with the same diagnostic objective, the results obtained showed that the volumetric dose index (CTDIvol) presented a higher dose value in the 12 cm diameter phantom. This phantom has smaller volume than the standard head phantom.
Downloads
References
ABURJAILE, W. N.; MOURAO, A. P. Development of a chest phantom for testing in Computed Tomography scans. Radiation Physics and Chemistry, v. 140, p. 275-277, 2017. https://doi.org/10.1016/j.radphyschem.2017.01.031
GOO, Hyun Woo. CT radiation dose optimization and estimation: an update for radiologists. Korean journal of radiology, v. 13, n. 1, p. 1-11, 2012. https://doi.org/10.3348/kjr.2012.13.1.1
UNSCEAR - United Nations Scientific Committee on the Effects of Atomic Radiation et al. Sources, Effects and Risks of Ionizing Radiation, United Nations Scientific Committee on the Effects of Atomic Radiation (UNSCEAR) 2020/2021 Report, Volume I: Report to the General Assembly, with Scientific Annex A-Evaluation of Medical Exposure to Ionizing Radiation. 2022. 354p. https://www.unscear.org/unscear/en/publications/2020_2021_1.html
MATHEWS, John D. et al. Cancer risk in 680 000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. Bmj, v. 346, p. 1-18, 2013. https://doi.org/10.1136/bmj.f2360
MOURÃO, Arnaldo Prata. Tomografia computadorizada: tecnologias e aplicações. Difusão Editora, 2018.
Mourão, A. P. et al. Dosimetry and Protocol Optimization of Computed Tomography Scans using Adult Chest Phantoms. Int J Radiol Imaging Technol, v. 5, p. 1-6, 2019. https://clinmedjournals.org/articles/ijrit/international-journal-of-radiology-and-imaging technology-ijrit-5-050.php?jid=ijrit
FINATTO, Jerusa et al. Estudo comparativo de descritor de dose em exames pediátricos de tomografia computadorizada. Brazilian Journal of Radiation Sciences, v. 3, n. 1A, p. 24-32, 2015. https://doi.org/10.15392/bjrs.v3i1A.126
IAEA - INTERNATIONAL ATOMIC ENERGY AGENCY. Dosimetry in diagnostic radiology: an international code of practice. Technical Reports Series No. 457, Viena, 2007.
BOLUS, Norman E. NCRP report 160 and what it means for medical imaging and nuclear medicine. Journal of Nuclear Medicine Technology, v. 41, n. 4, p. 255-260, 2013. https://doi.org/10.2967/jnmt.113.128728
YANG, Ching-Ching. Evaluation of impact of factors affecting CT radiation dose for optimizing patient dose levels. Diagnostics, v. 10, n. 10, p. 787, 2020. https://doi.org/10.3390/diagnostics10100787
ALONSO, Thêssa C.; MOURÃO FILHO, Arnaldo P.; DA SILVA, Teógenes A. Measurements of air kerma index in computed tomography: A comparison among methodologies. Applied Radiation and Isotopes, v. 138, p. 10-13, 2018. https://doi.org/10.1016/j.apradiso.2017.10.008
AAPM - American Association of Physicists in Medicine. Comprehensive methodology for the evaluation of radiation dose in x‐ray computed tomography. Report of AAPM Task Group 111: The Future of CT Dosimetry, 2010.
MIGLIORETTI, Diana L. et al. The use of computed tomography in pediatrics and the associated radiation exposure and estimated cancer risk. JAMA pediatrics, v. 167, n. 8, p. 700-707, 2013. https://jamanetwork.com/journals/jamapediatrics/article-abstract/1696279
SPAMPINATO, Maria Vittoria et al. Radiation dose and image quality in pediatric head CT. Radiation Protection Dosimetry, v. 182, n. 3, p. 310-316, 2018. https://doi.org/10.1093/rpd/ncy066
GOMEZ, Alvaro Mauricio Ladino. Estudo de dosimetria e qualidade de imagem em varreduras de tomografia computadorizada de cabeça utilizando objeto simulador. Master degree, Universidade Federal de Minas Gerais, Minas Gerais, Brasil 2017. https://repositorio.ufmg.br/handle/1843/BUOSAUTLEJ
DIWAKAR, MANOJ; KUMAR, MANOJ. A review on CT image noise and its denoising. Biomedical Signal Processing and Control, v. 42, p. 73-88, 2018. https://doi.org/10.1016/j.bspc.2018.01.010
DANCE, D. R. et al. Diagnostic radiology physics. International Atomic Energy Agency, v. 299, 2014. http://www.iaea.org/books
VÁSQUEZ, A. C. et al. Comparison of two respiration monitoring systems for 4D imaging with a Siemens CT using a new dynamic breathing phantom. Physics in Medicine & Biology, v. 57, n.9,p.131,2012. https://iopscience.iop.org/article/10.1088/00319155/57/9/N131/meta?casa_token=gs1yPU3CkcAAAAA:Zig077YuUDeYdw1DbbPy_35wj2UziWAMpDWLCunORvvinPX9Hm-u8NBQdndr2sp-1wP6oCiKRw
MANSON, E. N.; FLETCHER, J. J.; AMAUSI, J. H. Variation of noise with some scanning parameters for image quality test in 128 slice computed tomography scanner using catphan 7000 phantom. Journal of radiology and diagnostic methods, v. 1, p. 102, 2018. www.scholarena.com
CANTERA, A. C.; TORTOSA, M. L. Calidad de imagen. In: Curso de Garantia de qualidad en TC. Tarragona, España, p. 8–12, 2003.
NIST - National Institute of Standard and Tecnology. Values of the mass attenuation. Available at: < http://physics.nist.gov/PhysRefData/XrayMassCoef/tab4.html >. Last accessed: 08 Sept. 2020.
ICRP- International Commission on Radiological Protection. Diagnostic reference levels in medical imaging: review and additional advice. Oxford, p. 14. 2002. https://www.icrp.org/docs/DRL_for_web.pdf
Huda W. Medical Radiation Dosimetry. RSNA categorical course in diagnostic radiology physics: from invisible to visible—the science and practice of x-ray imaging and radiation dose optimization. RSNA., 29–39. 2006.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2023 Fernanda Stephanie Santos
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
The submission of original articles to the Brazilian Journal of Medical Physics implies the transfer, by the authors, of the rights of print and digital publication. Copyright for published articles remains with the author, with journal rights on first publication. Authors may only use the same results in other publications by clearly indicating this journal as the original publisher. As we are an open access journal, free use of articles in educational, scientific, non-commercial applications is allowed, as long as the source is cited.
The Brazilian Journal of Medical Physics is under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License (CC BY-NC-ND 4.0).