Study of CT Acquisition Protocols Using Two Head Phantoms

Authors

DOI:

https://doi.org/10.29384/rbfm.2023.v17.19849001706

Keywords:

Computed Tomography, Phantom, Dosimetry, Image Quality

Abstract

Computed Tomography (CT) scans promote a higher dose deposition than conventional radiology exams. These exams have significantly increased patient and collective doses and have become a global public health concern. There is a great need to improve protocols to seek for lower doses while maintaining the diagnostic image quality. The development of phantoms allows the testing of different acquisition protocols. In this study were tested two cylindrical head phantoms of polymethylmethacrylate (PMMA). One CT head phantom is the head standard test with 16 cm in diameter and the other head phantom developed is smaller at 12 cm in diameter. Both phantoms are 15 cm long. Different acquisition protocols were performed on a Philips CT scanner, Access model with 16 channels. The central slice of the phantoms was irradiated successively and measurements were performed using a pencil ionization chamber to obtain the CT air Kerma indexes in PMMA (Ck,PMMA,100) and CT dose indexes (CTDI).  From these results, the CT Dose Index values weighted and volumetric (CTDIw, CTDIvol) were obtained to 10 cm scans of the central region of the head phantoms, in helical mode. The scans were performed using different voltage values (80, 100 and 120 kV) and charge (mA.s). Dose values varied from 5.59 to 21.51 mGy. The highest recorded dose value was 21.51 mGy for the smaller head phantom and 19.25 mGy for the standard head phantom with 120 kV. Considering the generation of images with the same diagnostic objective, the results obtained showed that the volumetric dose index (CTDIvol) presented a higher dose value in the 12 cm diameter phantom. This phantom has smaller volume than the standard head phantom.

Downloads

Download data is not yet available.

References

ABURJAILE, W. N.; MOURAO, A. P. Development of a chest phantom for testing in Computed Tomography scans. Radiation Physics and Chemistry, v. 140, p. 275-277, 2017. https://doi.org/10.1016/j.radphyschem.2017.01.031

GOO, Hyun Woo. CT radiation dose optimization and estimation: an update for radiologists. Korean journal of radiology, v. 13, n. 1, p. 1-11, 2012. https://doi.org/10.3348/kjr.2012.13.1.1

UNSCEAR - United Nations Scientific Committee on the Effects of Atomic Radiation et al. Sources, Effects and Risks of Ionizing Radiation, United Nations Scientific Committee on the Effects of Atomic Radiation (UNSCEAR) 2020/2021 Report, Volume I: Report to the General Assembly, with Scientific Annex A-Evaluation of Medical Exposure to Ionizing Radiation. 2022. 354p. https://www.unscear.org/unscear/en/publications/2020_2021_1.html

MATHEWS, John D. et al. Cancer risk in 680 000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. Bmj, v. 346, p. 1-18, 2013. https://doi.org/10.1136/bmj.f2360

MOURÃO, Arnaldo Prata. Tomografia computadorizada: tecnologias e aplicações. Difusão Editora, 2018.

Mourão, A. P. et al. Dosimetry and Protocol Optimization of Computed Tomography Scans using Adult Chest Phantoms. Int J Radiol Imaging Technol, v. 5, p. 1-6, 2019. https://clinmedjournals.org/articles/ijrit/international-journal-of-radiology-and-imaging technology-ijrit-5-050.php?jid=ijrit

FINATTO, Jerusa et al. Estudo comparativo de descritor de dose em exames pediátricos de tomografia computadorizada. Brazilian Journal of Radiation Sciences, v. 3, n. 1A, p. 24-32, 2015. https://doi.org/10.15392/bjrs.v3i1A.126

IAEA - INTERNATIONAL ATOMIC ENERGY AGENCY. Dosimetry in diagnostic radiology: an international code of practice. Technical Reports Series No. 457, Viena, 2007.

BOLUS, Norman E. NCRP report 160 and what it means for medical imaging and nuclear medicine. Journal of Nuclear Medicine Technology, v. 41, n. 4, p. 255-260, 2013. https://doi.org/10.2967/jnmt.113.128728

YANG, Ching-Ching. Evaluation of impact of factors affecting CT radiation dose for optimizing patient dose levels. Diagnostics, v. 10, n. 10, p. 787, 2020. https://doi.org/10.3390/diagnostics10100787

ALONSO, Thêssa C.; MOURÃO FILHO, Arnaldo P.; DA SILVA, Teógenes A. Measurements of air kerma index in computed tomography: A comparison among methodologies. Applied Radiation and Isotopes, v. 138, p. 10-13, 2018. https://doi.org/10.1016/j.apradiso.2017.10.008

AAPM - American Association of Physicists in Medicine. Comprehensive methodology for the evaluation of radiation dose in x‐ray computed tomography. Report of AAPM Task Group 111: The Future of CT Dosimetry, 2010.

MIGLIORETTI, Diana L. et al. The use of computed tomography in pediatrics and the associated radiation exposure and estimated cancer risk. JAMA pediatrics, v. 167, n. 8, p. 700-707, 2013. https://jamanetwork.com/journals/jamapediatrics/article-abstract/1696279

SPAMPINATO, Maria Vittoria et al. Radiation dose and image quality in pediatric head CT. Radiation Protection Dosimetry, v. 182, n. 3, p. 310-316, 2018. https://doi.org/10.1093/rpd/ncy066

GOMEZ, Alvaro Mauricio Ladino. Estudo de dosimetria e qualidade de imagem em varreduras de tomografia computadorizada de cabeça utilizando objeto simulador. Master degree, Universidade Federal de Minas Gerais, Minas Gerais, Brasil 2017. https://repositorio.ufmg.br/handle/1843/BUOSAUTLEJ

DIWAKAR, MANOJ; KUMAR, MANOJ. A review on CT image noise and its denoising. Biomedical Signal Processing and Control, v. 42, p. 73-88, 2018. https://doi.org/10.1016/j.bspc.2018.01.010

DANCE, D. R. et al. Diagnostic radiology physics. International Atomic Energy Agency, v. 299, 2014. http://www.iaea.org/books

VÁSQUEZ, A. C. et al. Comparison of two respiration monitoring systems for 4D imaging with a Siemens CT using a new dynamic breathing phantom. Physics in Medicine & Biology, v. 57, n.9,p.131,2012. https://iopscience.iop.org/article/10.1088/00319155/57/9/N131/meta?casa_token=gs1yPU3CkcAAAAA:Zig077YuUDeYdw1DbbPy_35wj2UziWAMpDWLCunORvvinPX9Hm-u8NBQdndr2sp-1wP6oCiKRw

MANSON, E. N.; FLETCHER, J. J.; AMAUSI, J. H. Variation of noise with some scanning parameters for image quality test in 128 slice computed tomography scanner using catphan 7000 phantom. Journal of radiology and diagnostic methods, v. 1, p. 102, 2018. www.scholarena.com

CANTERA, A. C.; TORTOSA, M. L. Calidad de imagen. In: Curso de Garantia de qualidad en TC. Tarragona, España, p. 8–12, 2003.

NIST - National Institute of Standard and Tecnology. Values of the mass attenuation. Available at: < http://physics.nist.gov/PhysRefData/XrayMassCoef/tab4.html >. Last accessed: 08 Sept. 2020.

ICRP- International Commission on Radiological Protection. Diagnostic reference levels in medical imaging: review and additional advice. Oxford, p. 14. 2002. https://www.icrp.org/docs/DRL_for_web.pdf

Huda W. Medical Radiation Dosimetry. RSNA categorical course in diagnostic radiology physics: from invisible to visible—the science and practice of x-ray imaging and radiation dose optimization. RSNA., 29–39. 2006.

Published

2023-06-12

How to Cite

Stephanie Santos, F. (2023). Study of CT Acquisition Protocols Using Two Head Phantoms. Brazilian Journal of Medical Physics, 17, 706. https://doi.org/10.29384/rbfm.2023.v17.19849001706

Issue

Section

Artigo Original

Similar Articles

You may also start an advanced similarity search for this article.