How reliable are the ultrasonic output power values emitted by therapeutic ultrasound devices in physiotherapy?

Authors

  • Abraão da Rocha Melo Nascimento Brazilian Air Force Academy
  • Mariane Goulart Mocker Pinto
  • Thaís Pinório Omena
  • Wagner Coelho de Albuquerque Pereira
  • José Francisco Silva Costa-Júnior Brazilian Air Force Academy

DOI:

https://doi.org/10.29384/rbfm.2023.v17.19849001709

Keywords:

Therapeutic ultrasound, power, accuracy, maintenance, physiotherapy

Abstract

Therapeutic ultrasound is one of the most used electrophysical agents in physiotherapy. Despite this important and widespread application, therapeutic ultrasound devices (TUSD) may operate outside the range recommended by the International Standard IEC 61689:2013, and, thus, may not contribute to the treatment (if the ultrasonic output power, UOP, used in clinical practice are lower than the nominal power, NP) or even aggravate the pre-existing injury (if the UOP used is higher than the NP). The aim of this study was to verify whether the UOP values emitted by three TUSD from the Physiotherapy Section of the Esquadrão de Saúde de Pirassununga (named as TUSD1, TUSD2 and TUSD3) were in accordance with the IEC 61689:2013. Additionally, statistical tests were employed to assess the repeatability of UOP measurements made with the TUSD on three different days. An acoustic radiation force balance was used to measure the UOP values emitted by TUSD in three days of testing by each equipment, which were configured to operate in continuous mode. In addition, 10 UOP measurements were performed for each NP value with all available frequencies and effective radiation area. TUSD, even with up-to-date maintenance, presented several values of UOP with relative error greater than that preconized by IEC 61689:2013 (81.39% of 1800 UOP measurements with TUSD1, 68.61% of 1800 UOP measurements with TUSD2 and 75.33% of 1200 UOP measurements with TUSD3). So, It is important that the professional responsible for monitoring the maintenance/calibration of this equipment is trained to interpret the reports prepared by companies that provide maintenance/calibration services.

Downloads

Download data is not yet available.

Author Biography

José Francisco Silva Costa-Júnior, Brazilian Air Force Academy

José Francisco S. Costa-Júnior received the B.S. degree in Physics from Universidade Estadual de Santa Cruz, Ilhéus-BA, Brazil, in 2007, and the M.S. and Ph.D. degrees in n Biomedical Engineering from Federal University of Rio de Janeiro, Rio de Janeiro-RJ, Brazil, in 2010 and 2014, respectively. During his graduate studies, he worked on projects related to ultrasonic elastography, signal processing, metrology, physiotherapeutic ultrasound, phantoms, blood coagulation and characterization of biological tissues and phantoms. He is currently a Lieutenant in the Brazilian Air Force Academy, where he teaches subjects related to the field of physics, such as applied physics, electricity, mechanics and communication systems. His current research interests are ultrasonic signal and image processing, ultrasound radiation-force-based elastography, vibro-acoustography, conventional elastography, impulsive acoustic radiation force, bio-instrumentation and tissue characterization using ultrasound.

He is a member of the Brazilian Association of Biomedical Engineering.

References

Prentice WE. Therapeutic Modalities for Physical Therapists. New York: McGraw-Hill Education; 2001.

Kesiktas N, Karakas S, Gun K, Gun N, Murat S, Uludag M. Balneotherapy for chronic low back pain: a randomized, controlled study. Rheumatol Int. 2012; 32: 3193–99. http://dx.doi.org/10.1007/s00296-011-2163-9

Ricci AN, Dias KNC, Driusso P. The use of electrothermal and phototherapeutic methods for the treatment of fibromyalgia syndrome: a systematic review. Braz J Phys The. 2010; 14:1-9. https://doi.org/10.1590/S1413-35552010000100002

Silva FP, Severo-Silveira L, Plentz RDM, Durigan JLQ, Baroni BM.Electrophysical agents in clinical practice of orthopedic and sports physical therapists in Brazil. Fisioter. Pesqui. 2020; 27: 202-9. https://doi.org/10.1590/1809-2950/19019727022020

Agne JE. Eu sei eletroterapia. São Paulo: Pallotti; 2009..

IEC 61689:2013. Ultrasonics – physiotherapy systems – field specifications and methods of measurements in the frequency range 0,5 MHz to 5 MHz. International Electrotechnical Commission (IEC). 2013.

Merrick MA, Bernard KD, Devor ST, Williams MJ. Identical 3-MHz ultrasound treatments with different devices produce different intramuscular temperatures. J Orthop Sports Phys Ther. 2003; 33: 379-85.

https://doi.org/10.2519/jospt.2003.33.7.379

Johns LD, Demchak TJ, Straub SJ, Howard SM. The role of quantitative Schlieren assessment of physiotherapy ultrasound fields in describing variations between tissue heating rates of different transducers. Ultrasound Med. Biol. 2007; 33: 1911-17. https://doi.org/10.1016/j.ultrasmedbio.2007.06.012

Straub SJ, Johns LD, Howard SM. Variability in effective radiation area at 1 MHz affects ultrasound treatment intensity. Phys. Ther. 2008; 88: 50-7.

https://doi.org/10.2522/ptj.20060358

Baker KG, Robertson VJ, Duck FA. A review of therapeutic ultrasound: Biophysical effects. Phys Ther. 2001; 81: 1351–58.

Ishikawa NM, Alvarenga AV, Paes LFC, Pereira WCA, Machado JC. Análise do Desempenho de Equipamentos de Ultra-Som para Fisioterapia, operando na Cidade do Rio de Janeiro, Conforme a Norma NBR/IEC 1689 da ABNT. Rev Bras Fisioter. 2002; 6: 63-9.

Pye SD, Milford C. The performance of ultrasound physiotherapy machines in Lothian Region, Scotland, 1992, Ultrasound Med Biol. 1994; 20: 347-59.

https://doi.org/10.1016/0301-5629(94)90003-5

Guirro R, Santos SCB. Evaluation of the acoustic intensity of new ultrasound therapy equipment. Ultrasonics. 2002; 39: 553-7. https://doi.org/10.1016/s0041-624x(02)00251-2

Schabrun S, Walker H, Chipchase L. How accurate are therapeutic ultrasound machines? Hong Kong Physiother J. 2008; 26: 39-44.

https://doi.org/10.1016/S1013-7025(09)70006-5

Shaw A, Hodnett M. Calibration and measurement issues for therapeutic ultrasound. Utrasonics. 2008; 48: 234-52.

https://doi.org/10.1016/j.ultras.2007.10.010

Inmetro, Vocabulário Internacional de Metrologia (VIM 2012): Conceitos Fundamentais e termos associados. Brasília: INMETRO; 2012.

Werkema C. Avaliação de sistemas de medição. Rio de Janeiro: Elsevier Brasil; 2013.

Razali NM, Wah YB. Power comparisons of Shapiro-Wilk, Kolmogorov-Smirnov, Lilliefors and Anderson-Darling tests. J. Stat. Model. 2011; 2: 21-33.

Ferrari CB, Andrade MAB, Adamowski JC, Guirro RRJ. Evaluation of Therapeutic Ultrasound Equipments Performance. Ultrasonics. 2010; 50: 704-9.

https://doi.org/10.1016/j.ultras.2010.02.006

Durando G, Guglielmone C. Ultrasound physiotherapy devices: how to measure them. IEEE Instrum. Meas Mag. 2016; 19: 15-48. https://doi.org/10.1109/MIM.2016.7579065.

Published

2023-07-17

How to Cite

da Rocha Melo Nascimento, A., Goulart Mocker Pinto, M., Pinório Omena, T., Coelho de Albuquerque Pereira, W., & Silva Costa-Júnior, J. F. (2023). How reliable are the ultrasonic output power values emitted by therapeutic ultrasound devices in physiotherapy?. Brazilian Journal of Medical Physics, 17, 709. https://doi.org/10.29384/rbfm.2023.v17.19849001709

Issue

Section

Artigo Original