CT dose estimations using the ICRU/AAPM phantom
DOI:
https://doi.org/10.29384/rbfm.2022.v16.19849001685Keywords:
tomografia computadorizada, dosimetria, dose de equilíbrio, phantomAbstract
A proposed method to obtain the dose for Computed Tomography was evaluated on long a phantom, made of polyethylene. This phantom is 600 mm large. Adopting an empirical characterized function , a tendency of the asymptotical grow of the dose by the increase of the scanning length can estimate the dose in an infinite scanning length for a single axial rotation of the x-ray tube. The curve of was estimated for different clinical protocols: tension, pitch, head/body, and center/peripheral measurements. The ultrahigh molecular weight polyethylene phantom of 600 mm length, named ICRU/AAPM phantom, was proposed by a partnership between ICRU and AAPM (Task Group 200). This length provides a higher scanner length then PMMA phantom for , whose measurement is 100 mm. The measures of ICRU/AAPM phantom were performed on General Electric, Toshiba, and Philips scanners. The measurement was performed on a thimble ion chamber inserted in the center or periphery of the phantom. Using the measurements, functions (equilibrium dose, equilibrium length and scatter fraction) were obtained. A single scan has been demonstrated to be sufficient to obtain the function.
Downloads
References
Dixon RL. The Physics of CT Dosimetry: CTDI and Beyond. 1st ed. Boca Raton: CRC Press; 2019.
AAPM. The Measurement, Reporting, and Management of Radiation Dose in CT. American Association of Physicists in Medicine. 2008;AAPM Report 096.
Boone JM. The trouble with CTDI100. Medical Physics. 2007;34(4):1364-71.
AAPM. Comprehensive Methodology for the Evaluation of Radiation Dose in X-Ray Computed Tomography. American Association of Physicists in Medicine. 2010;AAPM Report 111.
ICRU. Radiation Dosimetry and Image Quality Assessment in Computed Tomography. International Commission on Radiation Units and Measurements. 2012;Report 87.
AAPM. The Design and Use of the ICRU/AAPM CT Radiation Dosimetry Phantom: An Implementation of AAPM Report 111. American Association of Physicists in Medicine. 2020;AAPM Report 200.
Costa PR, Nersissian DY, Umisedo NK, Gonzales AHL, Fernandez-Varea JM. A comprehensive Monte Carlo study of CT dose metrics proposed by the AAPM Reports 111 and 200. Medical Physics. 2022;49(1):201-18.
Dixon RL, Ballard AC. Experimental validation of a versatile system of CT dosimetry using a conventional ion chamber: beyond CTDI100. Medical physics. 2007;34(8).
Radcal Corporation. 10X6-0.6C: The Chamber for Computed Tomography Dose Index (CTDI); 2011.
Leitz W, Axelsson B, Szendro G. COMPUTED-TOMOGRAPHY DOSE ASSESSMENT - A PRACTICAL APPROACH. Radiation Protection Dosimetry. 1995;57(1-4):377-80.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2022 Yuri Stefani, Denise Yanikian Nersissian, Paulo Roberto Costa
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
The submission of original articles to the Brazilian Journal of Medical Physics implies the transfer, by the authors, of the rights of print and digital publication. Copyright for published articles remains with the author, with journal rights on first publication. Authors may only use the same results in other publications by clearly indicating this journal as the original publisher. As we are an open access journal, free use of articles in educational, scientific, non-commercial applications is allowed, as long as the source is cited.
The Brazilian Journal of Medical Physics is under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License (CC BY-NC-ND 4.0).